Megger 220123-47 120kV DC Dielectric Test Set
Specifications :
Input Power: 220/240 Vac, 50/60 Hz.
Operates Like a Full-Wave Rectified Unit (Filtered Half-Wave Rectification).
Allows for a simple circuit scheme for long-term reliability.
Lightweight High-Voltage Module.
Test Voltage: 0 - 120kV DC.
Max Power System Voltage (Phase to Phase): 35kV AC.
Output Current: 5 mA for 5 min; 2 mA continuous.
Display: Digital.
Ammeter Ranges: 0 to 19.9 µA, 0 to 199 µA, 0 to 1.99 mA, 0 to 5 mA.
Ammeter Resolution: To 0.1 µA on lowest range.
Voltmeter Resolution: To 100 V over entire range.
Ripple: Less than 2% on capacitive samples at continuous rated output.
Input-supply-line circuit breaker.
Output current overload relay.
Zero-start interlock for high-voltage output.